RecallWatchUSA

Home / Recalls / Medical Devices

Conmed Corporation recalls Adult Dual Dispersive Electrodes (for patients >15 kg.), use with the Birtcher Pad Sensing System (PSS) on the Birtcher

Recalled August 31, 2016 · FDA recall Z-2598-2016 · Conmed

Hazard
A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn. · FDA Class II
Units
124,333 units
Sold at
Worldwide Distribution -- USA, including AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & Puerto Rico; and, the countries of Australia, Canada, Belgium, Brunei Darussalam, China, France, Guatemala, Hong Kong, Israel, Italy, Japan, South Korea, Malaysia, Mexico, New Zealand, Norway, Portugal, Saudi Arabia, Spain, Sweden, and United Arab Emirates.

Description

A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.

Product
Adult Dual Dispersive Electrodes (for patients >15 kg.), use with the Birtcher Pad Sensing System (PSS) on the Birtcher Systems 6000, 6400, 6500 and CONMED Systems 7500 and 7550 ABC¿ generators or conventional electrosurgery units with similar contact monitoring systems, 10' (3.05m) Cable, Catalog Number 7-382. Intended to be used for the dispersion and return to the electrosurgical generator.
Manufacturer(s)
Conmed Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Conmed Corporation recalls Adult Dual Dispersive Electrodes (for patients >15 kg.), use with the Birtcher Pad Sensing System (PSS) on the Birtcher | RecallWatchUSA