RecallWatchUSA

Home / Recalls / Medical Devices

ConMed Corporation recalls Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurg

Recalled April 19, 2017 · FDA recall Z-1758-2017 · ConMed

Hazard
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile. · FDA Class II
Units
N/A
Sold at
Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.

Description

For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.

Product
Sterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons , 220 VOLTS AC, REF/Catalog Number 7-900-220, Rx Only, There is no problem with the Hyfrecator 2000 Electrosurgical Units --- ASSEMBLED IN MEXICO -- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502 ---
Manufacturer(s)
ConMed Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov