Home / Recalls / Medical Devices
Cook Inc. recalls Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108 G14587 G15
Recalled February 12, 2020 · FDA recall Z-0966-2020 · Cook
Hazard
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip. · FDA Class II
Units
13388 total
Sold at
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Description
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Product
Angled Tip Ureteral Catheter Set with Bentson PTFE Wire Guide, Ureteral Catheter, Global Product No. G19108 G14587 G15302 G14326
Manufacturer(s)
Cook Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →