Home / Recalls / Medical Devices
Cook Inc. recalls MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidifie
Recalled April 25, 2018 · FDA recall Z-1378-2018 · Cook
Hazard
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater. · FDA Class II
Units
4,451
Sold at
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)
Description
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.
Product
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000) G20079 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Manufacturer(s)
Cook Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →