Cook Inc. recalls MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidi

Recalled April 25, 2018 · FDA recall Z-1377-2018 · Cook

Hazard
The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater. · FDA Class II
Units
483
Sold at
Nationwide (K-IMNC-100 -US G46022) . (K-IMNC-100 G20079 manufactured in Australia and distributed worldwide)

Description

The temperature of the incubation lids of the MINC Bench top Incubators may become too high or too low for proper gamete/embryo incubation due to fatigue failure of the flat flex lid heater.

Product
MINC Benchtop Incubator (Mini Incubator) catalog (K-MINC-1000 US) G46022 is a microprocessor controlled, gassed, humidified, incubator intended for use in cell culture.
Manufacturer(s)
Cook Inc.