Home / Recalls / Medical Devices
Cordis Corporation recalls Cordis 5F SUPER TORQUE PIG PIGTAIL SPECIAL, Angiographic Catheter, REF SRD7040MB
Recalled September 22, 2021 · FDA recall Z-2429-2021 · Cordis
Hazard
Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement. · FDA Class I
Units
1610 units
Sold at
Worldwide distribution - US Nationwide distribution and the countries of Canada, Europe.
Description
Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement.
Product
Cordis 5F SUPER TORQUE PIG PIGTAIL SPECIAL, Angiographic Catheter, REF SRD7040MB
Manufacturer(s)
Cordis Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →