Home / Recalls / Medical Devices
CORPAK MedSystems, Inc. recalls CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 12FR (4mm); Product Usage: CORFLO PEG Tubes are intende
Recalled May 31, 2017 · FDA recall Z-2161-2017 · CORPAK MedSystems
Hazard
Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress. · FDA Class II
Units
2 cases (5 units per case)
Sold at
Worldwide distribution in the states of NH, VT, CA and the countries of UK, Ireland, France, Austria, Norway, Sweden, New Zealand
Description
Firm is initiating voluntary correction of CORFLO PEG with ENFit. Barb connector (stem) of the adapter may break unexpectedly when exposed to stress.
Product
CORFLO, PEG Adaptor Repair Kit with ENFit Connectors, Size: 12FR (4mm); Product Usage: CORFLO PEG Tubes are intended for delivery of enteral nutrition, water and medications into the stomach. The PEG adapter is the connection point between the tubing and the feeding set/syringe. The adapter is designed as a replaceable component.
Manufacturer(s)
CORPAK MedSystems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →