Covidien recalls Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941

Recalled July 30, 2025 · FDA recall Z-2151-2025 · Covidien

Hazard
Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting · FDA Class II
Units
264,790
Sold at
Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bhutan, Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic Of, Ireland, Isle of Man, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, Zimbabwe.

Description

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting

Product
Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN
Manufacturer(s)
Covidien

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov