Home / Recalls / Medical Devices
Covidien Llc recalls Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code: 50514T - Product Usage: for Draina
Recalled May 20, 2020 · FDA recall Z-1984-2020 · Covidien
Hazard
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment · FDA Class II
Units
468 units
Sold at
US Nationwide distribution.
Description
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Product
Dover 100% Silicone Foley Catheter with Temperature Sensor 14FR 5CC2W, Product Code: 50514T - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Manufacturer(s)
Covidien Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →