RecallWatchUSA

Home / Recalls / Medical Devices

Covidien Llc recalls Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 2000 mL Dra

Recalled May 20, 2020 · FDA recall Z-1988-2020 · Covidien

Hazard
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment · FDA Class II
Units
3470 units
Sold at
US Nationwide distribution.

Description

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Product
Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P14TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Manufacturer(s)
Covidien Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov