RecallWatchUSA

Home / Recalls / Medical Devices

Covidien Llc recalls Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT -

Recalled May 20, 2020 · FDA recall Z-1987-2020 · Covidien

Hazard
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment · FDA Class II
Units
0
Sold at
US Nationwide distribution.

Description

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Product
Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Manufacturer(s)
Covidien Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Covidien Llc recalls Dover 100% Silicone Silver-Hydrogel Coated Foley Catheter with Temperature Sensor 18FR 5CCCC2W, Product Code: 50518ICT - | RecallWatchUSA