Home / Recalls / Medical Devices
Covidien Llc recalls Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-co
Recalled May 20, 2020 · FDA recall Z-1993-2020 · Covidien
Hazard
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment · FDA Class II
Units
120 units
Sold at
US Nationwide distribution.
Description
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Product
Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P18XTS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Manufacturer(s)
Covidien Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →