RecallWatchUSA

Home / Recalls / Medical Devices

Covidien, LP recalls Covidien Signia Grey Small Diameter Curved Tip Intelligent Reload 30 mm Vascular 8 mm - Short Item Code: SIGSDS30CTV

Recalled April 21, 2021 · FDA recall Z-1384-2021 · Covidien

Hazard
Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure. · FDA Class II
Units
11391 units WW: 5982 US 5409 OUS
Sold at
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom.

Description

Potential for the device safety interlock to fail. The safety interlock prevents an empty single-use reload from being erroneously fired a second time. If the safety interlock fails and the empty single-use reload is fired a second time, the stapler may cut tissue but no staples will deploy from the empty reload, potentially resulting in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open procedure.

Product
Covidien Signia Grey Small Diameter Curved Tip Intelligent Reload 30 mm Vascular 8 mm - Short Item Code: SIGSDS30CTV
Manufacturer(s)
Covidien, LP

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov