RecallWatchUSA

Home / Recalls / Medical Devices

Covidien, LP recalls Mild Chromic Gut Absorbable Sutures, CH GUT 5-0 12 UD SS-14 SS-14 (G-1766-K), CHROMIC GUT 6-0 45CM HE-1 X12 (G-1780K)

Recalled April 19, 2023 · FDA recall Z-1389-2023 · Covidien

Hazard
Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk. · FDA Class II
Units
960 units
Sold at
Worldwide Distribution: US (nationwide) to States of: AR, CA, FL, GA, LA, MA, NC, NY, OH, WA; and countries of: Canada, Colombia, El Salvador, Indonesia, Nepal, Puerto Rico, Taiwan, and Thailand.

Description

Medtronic was made aware of a supplier calibration issue during the contract sterilization process that may lead to unsterile product. Use of affected product may lead to an unspecified infection or, in the case of sutures used in the eye, pose a loss of vision risk.

Product
Mild Chromic Gut Absorbable Sutures, CH GUT 5-0 12 UD SS-14 SS-14 (G-1766-K), CHROMIC GUT 6-0 45CM HE-1 X12 (G-1780K)
Manufacturer(s)
Covidien, LP

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Covidien, LP recalls Mild Chromic Gut Absorbable Sutures, CH GUT 5-0 12 UD SS-14 SS-14 (G-1766-K), CHROMIC GUT 6-0 45CM HE-1 X12 (G-1780K) | RecallWatchUSA