RecallWatchUSA

Home / Recalls / Medical Devices

Covidien LP recalls Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 888

Recalled February 7, 2024 · FDA recall Z-0844-2024 · Covidien

Hazard
Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level. · FDA Class II
Units
46,812 US; 68,976 OUS
Sold at
Worldwide US Nationwide distribution including in the states of AL, AZ, CA, FL, GA, HI, IA, IL, IN, MD, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, PR, TN, TX, VA, WA, WV and the countries of Australia, Bahrain, Belgium, Bulgaria, Canary Islands, China, Cyprus, Czech Republic, Finland, France, Germany, Greece, Hong Kong, Hungary, Italy, Japan, Korea, Republic Of Kosovo, Mongolia, Netherlands, New Zealand, Panama, Poland, Portugal, Puerto Rico, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Kingdom.

Description

Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.

Product
Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 88861919-51 SURGILON* 2-0 BLK 7X75CMPCT 88861919-71 SURGILON* 1 BLK 7X75CM PCT
Manufacturer(s)
Covidien LP

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Covidien LP recalls Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 888 | RecallWatchUSA