Covidien LP recalls VersaOne Reusable Positioning Cannula 11mm Standard, Model Number RC11STS
Recalled May 3, 2023 · FDA recall Z-1464-2023 · Covidien
Hazard
There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together. · FDA Class II
Units
51 units
Sold at
Worldwide - US Nationwide distribution in the states of NC and the countries of Belgium, Denmark, France, Germany, India, Italy, Netherlands, Spain, Sweden.
Description
There is a manufacturing non-conformance that may result in a metal shaving (burr) on the inside of the VersaOneTM Reusable Positioning Cannula, located where the Cannula and Body components are welded together.
Product
VersaOne Reusable Positioning Cannula 11mm Standard, Model Number RC11STS
Manufacturer(s)
Covidien LP
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →