RecallWatchUSA

Home / Recalls / Medical Devices

Crossroads Extremity Systems Llc recalls MotoCLIP(TM) Super Elastic Fusion System, Size 18mm x 14mm x 14mm, REF 1442-1814, Rx Only, STERILE R.

Recalled March 29, 2017 · FDA recall Z-1487-2017 · Crossroads Extremity Systems

Hazard
Seven 18mm x 18mm x 18 mm implants had been placed in kits, packed and labeled as 18mm x 14mm x 14mm. · FDA Class II
Units
20 units
Sold at
Al, FL, IN, Il, NE, TX, NM, CO, AZ

Description

Seven 18mm x 18mm x 18 mm implants had been placed in kits, packed and labeled as 18mm x 14mm x 14mm.

Product
MotoCLIP(TM) Super Elastic Fusion System, Size 18mm x 14mm x 14mm, REF 1442-1814, Rx Only, STERILE R.
Manufacturer(s)
Crossroads Extremity Systems Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov