CSL Behring GmbH recalls Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Sie

Recalled April 29, 2020 · FDA recall Z-1790-2020 · CSL Behring

Hazard
Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring · FDA Class II
Units
297 PCK US; 1247 PCK OUS
Sold at
IN, KS, MO, OH, VA Foreign: Australia, Austria, Bangladesh, Belgium, Canada, China Czech Republic, Egypt, Finland, France, Germany, Great Britain, Hungary India, Italy, Japan, Mexico, Netherlands, Norway, Poland, Portugal, Republic Korea, Russia Federation, Slovakia, Slovenia, South Africa, Spain Sweden, Switzerland, Taiwan

Description

Negative bias for the carbohydrate-deficient transferrin (CDT) measurement in comparison to the HPLC method, observed to occur after 15 months based on shelf life monitoring

Product
Siemens N Latex CDT Kit- IVD quantitative determination of carbohydrate-deficient transferrin (CDT) in human serum. Siemens Material Number (SMN): 10445997
Manufacturer(s)
CSL Behring GmbH
Made in
Germany

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov