RecallWatchUSA

Home / Recalls / Medical Devices

Custom Spine, Inc. recalls Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus

Recalled August 7, 2013 · FDA recall Z-1819-2013 · Custom Spine

Hazard
On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm. · FDA Class II
Units
18 units
Sold at
Nationwide Distribution including Nebraska and Virginia.

Description

On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw size. The product is laser marked 8.5mm X 45mm, but its length measured 40mm.

Product
Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus Dr. Parsippany, NJ 07054 USA. TEL: 973-808-0019 The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the lumbar and/or sacral spine for pedicular and non-pedicular fixation.
Manufacturer(s)
Custom Spine, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Custom Spine, Inc. recalls Custom Spine ISSYS LP Inverted Screw System. 8.5MM X 45 MM ISSYS LP Screw Rx Non-Sterile. Custom Spine, Inc. 9 Campus | RecallWatchUSA