RecallWatchUSA

Home / Recalls / Medical Devices

Cutera, Inc. recalls Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product is manufactur

Recalled December 18, 2013 · FDA recall Z-0427-2014 · Cutera

Hazard
Cutera received two adverse events associated with the edge of the electrode, when it was pressed against neck skin over the thyroid cartilage and due to the absence of adequate underlying conductive tissue, increased skin heating occurred. Both patients developed linear welts and blisters consistent with partial thickness thermal burns to the skin. · FDA Class II
Units
451
Sold at
Worldwide Distribution - USA (nationwide) and Internationally to Canada

Description

Cutera received two adverse events associated with the edge of the electrode, when it was pressed against neck skin over the thyroid cartilage and due to the absence of adequate underlying conductive tissue, increased skin heating occurred. Both patients developed linear welts and blisters consistent with partial thickness thermal burns to the skin.

Product
Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product is manufactured and distributed by Cutera, Inc., Brisbane, CA The truSculpt massage device is intended to provide a temporary reduction in the appearance of cellulite. The truSculpt infrared (optional) and RF energies are intended to provide topical heating for the purpose of elevating tissue temperature for the treatment of selected medical conditions such as relief of pain, muscle spasms, and increase in local circulation.
Manufacturer(s)
Cutera, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Cutera, Inc. recalls Cutera brand truSculpt infrared System (optional), Massager, Vacuum, radio frequency induced heat, Product is manufactur | RecallWatchUSA