RecallWatchUSA

Home / Recalls / Medical Devices

Cytocell Ltd. recalls Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists

Recalled February 17, 2021 · FDA recall Z-1099-2021 · Cytocell

Hazard
Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green. · FDA Class II
Units
72
Sold at
Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada.

Description

Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green.

Product
Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric) to the CRLF2 gene
Manufacturer(s)
Cytocell Ltd.
Made in
United Kingdom

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov