Cytocell Ltd. recalls Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML o

Recalled March 16, 2022 · FDA recall Z-0741-2022 · Cytocell

Hazard
individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result · FDA Class II
Units
19 kits
Sold at
US Nationwide distribution in the states of Hawaii, Missouri, Pennsylvania, Tennessee.

Description

individual components have been labelled with incorrect colours. The red and green colours are opposite to those specified in the product labelling. The EGR1 probe is labelled in green rather than red and the control probe labelled in red rather than green. The expected positive signal pattern in the case of 1x deleted 5q will be 1G2R instead of 1R2G, an incorrect result

Product
Del(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH024
Manufacturer(s)
Cytocell Ltd.
Made in
United Kingdom