Home / Recalls / Medical Devices
Cytrellis Biosystems, Inc. recalls Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of mo
Recalled June 7, 2023 · FDA recall Z-1648-2023 · Cytrellis Biosystems
Hazard
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring. · FDA Class II
Units
89 units
Sold at
Nationwide
Description
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.
Product
Cytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and severe wrinkles in the mid and lower face. Model #: Gen 4
Manufacturer(s)
Cytrellis Biosystems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →