RecallWatchUSA

Home / Recalls / Medical Devices

Daavlin Distributing Company recalls Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders

Recalled July 22, 2026 · FDA recall Z-2742-2026 · Daavlin Distributing

Hazard
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000). · FDA Class II
Units
2 units
Sold at
US Nationwide distribution.

Description

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

Product
Photera 400, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Manufacturer(s)
Daavlin Distributing Company

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov