Home / Recalls / Medical Devices
Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0
Recalled March 15, 2023 · FDA recall Z-1231-2023 · Datascope
Hazard
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted. · FDA Class II
Units
11,792 units (5.408 US, 6,384 OUS)
Sold at
Nationwide US distribution. Worldwide international distribution.
Description
Firm has received reports of damaged, worn, or torn O-rings on the Cardiosave Pump Console quick disconnect fitting, which results in helium tank leaks. If the helium supply is depleted, therapy will be interrupted.
Product
Cardiosave Hybrid Intra-Aortic Balloon Pump. Model Numbers: D998-00-0800-31, D998-00-0800-32, D998-00-0800-33, D998-00-0800-35, D998-00-0800-36, D998-00-0800-45, D998-00-0800-52, D998-00-0800-53, D998-00-0800-55, D998-00-0800-65, D998-00-0800-83, D998-UC-0800-31, D998-UC-0800-33, D998-UC-0800-52, D998-UC-0800-53, D998-UC-0800-55.
Manufacturer(s)
Datascope Corp.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →