Datascope Corp. recalls Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbe
Recalled March 15, 2023 · FDA recall Z-1145-2023 · Datascope
Hazard
An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit. · FDA Class I
Units
4062 Total
Sold at
Worldwide Distribution: US (nationwide) and OUS (Foreign) ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BELARUS, BELGIUM, BRAZIL, BRUNEI , BULGARIA, CANADA, CHILE, CHINA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FINLAND, FRANCE, GERMANY, GHANA, HONG KONG, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PHILIPPINES, POLAND, PORTUGAL, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TURKEY, TURKMENISTAN, UNITED ARAB EMIRATES, UNITED KINGDOM, VIETNAM, AND YEMEN.
Description
An unexpected shutdown of the IABP may occur due to a failure of the connection between the Coiled Cord cable and the Cable Assembly backplane to the Coiled Cord cable which provides the communication between the display head and base unit.
Product
Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Containing Coiled Cord cable (part number 0012-00-1801, Model Numbers 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-00-0800-55
Manufacturer(s)
Datascope Corp.