Datascope Corp. recalls Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IA

Recalled October 27, 2021 · FDA recall Z-0091-2022 · Datascope

Hazard
A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers. · FDA Class I
Units
131 batteries, 11,517 pumps
Sold at
Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium, Canada, China, Czech Republic, Egypt, France, Germany, India, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, Pakistan, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, UK, United Arab Emirates, Philippines, and Russia

Description

A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

Product
Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below Model Numbers. 1. Cardiosave Hybrid. Model Numbers: 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55. 2. Cardiosave Rescue. Model Number: 0998-00-0800-83.
Manufacturer(s)
Datascope Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Datascope Corp. recalls Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IA | RecallWatchUSA