RecallWatchUSA

Home / Recalls / Medical Devices

Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85

Recalled October 6, 2021 · FDA recall Z-2616-2021 · Datascope

Hazard
The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions. · FDA Class II
Units
3969 total (US); 3550 total (OUS)
Sold at
Nationwide domestic distribution. Foreign distribution worldwide.

Description

The Cardiosave IABP may unexpectedly shut down when the device is running on AC power, only one battery is installed in the IABP, and the battery is physically removed while the battery is being charged. This occurs during a very specific set of conditions.

Product
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Nos. 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Manufacturer(s)
Datascope Corp.