Home / Recalls / Medical Devices
Datascope Corp. recalls Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 099
Recalled January 25, 2023 · FDA recall Z-0886-2023 · Datascope
Hazard
The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter. · FDA Class I
Units
46 Total
Sold at
Worldwide Distribution - US Nationwide and the countries of Albania, Angola, Argentina, Australia, Austria, AZERBAIJAN, BAHRAIN, BANGLADESH, Barbados, Belarus, Belgium, BOSNIA and HERZEGOVINA, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, C¿te d'Ivoire, Croatia, CZECH REPUBLIC, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Ghana, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, TRINIDAD AND TOBAGO, Tunisia, Turkey, Turkmenistan, Ukraine, UNITED ARAB EMIRATES, United Kingdom, Vietnam, Yemen, and Zambia.
Description
The Cardiosave IABP may shut down unexpectedly due to blood entering in the Cardiosave IABP when therapy is provided with a compromised intra-aortic balloon catheter.
Product
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85.
Manufacturer(s)
Datascope Corp.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →