RecallWatchUSA

Home / Recalls / Medical Devices

Datascope Corp. recalls MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta

Recalled August 2, 2023 · FDA recall Z-2219-2023 · Datascope

Hazard
During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death. · FDA Class II
Units
321,609 total kits
Sold at
Domestic distribution nationwide. International distribution worldwide.

Description

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Product
MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES D684-00-0295-01 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) D684-00-0295-02 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0295-03 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (CHINA) D684-00-0295-05 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES AND APA (CHINA) D684-00-0295-06 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA D684-00-0295-07 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES W/O STATLOCK (APA) - INDIA D684-00-0295-08 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (TURKEY) D684-00-0295-09 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES (APA) (TURKEY) D684-00-0295-10 MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES, US ONLY D684-00-0295-01U MEGA 7.5Fr. 40cc IAB WITH ACCESSORIES APA, US ONLY D684-00-0295-02U MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES D684-00-0294-01 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (APA) D684-00-0294-02 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (W/O STATLOCK) D684-00-0294-03 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (CHINA) D684-00-0294-05 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES AND APA (CHINA) D684-00-0294-06 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK - INDIA D684-00-0294-07 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES W/O STATLOCK (APA) INDIA D684-00-0294-08 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (TURKEY) D684-00-0294-09 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES (APA) (TURKEY) D684-00-0294-10 MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES, US ONLY D684-00-0294-01U MEGA 7.5Fr. 30cc IAB WITH ACCESSORIES APA, US ONLY D684-00-0294-02U PACKAGED INSERTION KIT - MEGA 7.5 Fr. 30/40 cc IABs D884-00-0019-21
Manufacturer(s)
Datascope Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Datascope Corp. recalls MEGA 7.5Fr.IAB , Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta | RecallWatchUSA