Datascope Corporation recalls BEQ-T 25502 1/4 Better Bladder , Custom Tubing Kit, Catalog No. 709000434

Recalled May 16, 2018 · FDA recall Z-1722-2018 · Datascope

Hazard
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow. · FDA Class II
Units
40
Sold at
Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.

Description

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Product
BEQ-T 25502 1/4 Better Bladder , Custom Tubing Kit, Catalog No. 709000434
Manufacturer(s)
Datascope Corporation