RecallWatchUSA

Home / Recalls / Medical Devices

Datascope Corporation recalls BEQ-TOP 33700 ECC 3/8" PACK, Custom Tubing Pack with Bioline Coating, Catalog No. 701053845

Recalled May 16, 2018 · FDA recall Z-1716-2018 · Datascope

Hazard
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow. · FDA Class II
Units
22
Sold at
Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.

Description

The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.

Product
BEQ-TOP 33700 ECC 3/8" PACK, Custom Tubing Pack with Bioline Coating, Catalog No. 701053845
Manufacturer(s)
Datascope Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Datascope Corporation recalls BEQ-TOP 33700 ECC 3/8" PACK, Custom Tubing Pack with Bioline Coating, Catalog No. 701053845 | RecallWatchUSA