DATEX--OHMEDA, INC. recalls GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator

Recalled October 23, 2024 · FDA recall Z-0048-2025 · DATEX--OHMEDA

Hazard
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose. · FDA Class I
Sold at
Worldwide distribution.

Description

GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.

Product
GE Heathcare Giraffe OmniBed Carestation, Model Number 2082844-001-XXX; neo natal incubator
Manufacturer(s)
DATEX--OHMEDA, INC.