RecallWatchUSA

Home / Recalls / Medical Devices

Datex-Ohmeda Inc recalls Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 20826

Recalled June 24, 2026 · FDA recall Z-2357-2026 · Datex-Ohmeda

Hazard
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient. · FDA Class I
Units
550 (367 US; 183 OUS)
Sold at
Pending

Description

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

Product
Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 2082663-001, 2082663-002, M1241424 Product impacted only if service blender M1091607-R was installed during the last Preventative Maintenance
Manufacturer(s)
Datex-Ohmeda Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov