Dental EZ Stardental Division recalls StarDental Classique(TM) Diamond Dental Bur. Used for specific applications in dental procedures.

Recalled June 19, 2013 · FDA recall Z-1522-2013 · Dental EZ Stardental Division

Hazard
Recall was initiated because a manufacturing issue was found that prevented the diamond coating from adhering properly to the shank of the diamond bur for these lots. · FDA Class II
Units
3,560
Sold at
Worldwide Distribution-USA (nationwide) including the states of CA, FL, IA, IN, NJ, NV, NY, PA, TN, TX, and WI, and the country of Japan.

Description

Recall was initiated because a manufacturing issue was found that prevented the diamond coating from adhering properly to the shank of the diamond bur for these lots.

Product
StarDental Classique(TM) Diamond Dental Bur. Used for specific applications in dental procedures.
Manufacturer(s)
Dental EZ Stardental Division