DePuy Orthopaedics, Inc. recalls GLOBAL¿ Anchor Peg Glenoid Plus (APG+) 2.5MM Breakaway Guide Pin Part # 2230-00-019, Non-Sterile, DePuy, Johnson & Joh

Recalled April 16, 2014 · FDA recall Z-1437-2014 · DePuy Orthopaedics

Hazard
Urgent Device Correction notifications are being sent to USA distributors with follow-up to surgeons concerning updated surgical techniques stressing the possibility of the shoulder pin guide breaking and potentially being left in the patient. · FDA Class II
Units
21,675
Sold at
USA Nationwide Distribution including Canada

Description

Urgent Device Correction notifications are being sent to USA distributors with follow-up to surgeons concerning updated surgical techniques stressing the possibility of the shoulder pin guide breaking and potentially being left in the patient.

Product
GLOBAL¿ Anchor Peg Glenoid Plus (APG+) 2.5MM Breakaway Guide Pin Part # 2230-00-019, Non-Sterile, DePuy, Johnson & Johnson Product Usage: Usage: The GLOBAL APG+ 2.5mm Guide Pin is a single-use instrument that is used to assist the surgeon in the orientation of the cannulated reamer(s) and cannulated central drills in the preparation of the glenoid surface. The GLOBAL APG+ 2.5mm Guide Pin is validated for use with the GLOBAL APG+, GLOBAL STEPTECH APG, and DELTA XTEND Systems. The guide pin is scored in three locations, which allows the surgeon to select a desired pin length. The single-use shoulder instrument is made from 316L Stainless Steel.
Manufacturer(s)
DePuy Orthopaedics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov