Home / Recalls / Medical Devices
DePuy Spine, Inc. recalls DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final t
Recalled January 23, 2013 · FDA recall Z-0685-2013 · DePuy Spine
Hazard
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification. · FDA Class II
Units
140 units
Sold at
Worldwide Distribution - USA (nationwide) and internationally to Belgium, Egypt, Ireland, Italy, Portugal, and the UK.
Description
A component within the handle of the Expedium Offset Torque Wrench is prematurely wearing, potentially resulting in the instrument recording torque values outside of the specification.
Product
DePuy Spine EXPEDIUM Offset Torque Wrench Product Code 2797-29-510 The device is intended for use during the final tightening of an Expedium Offset Spinal Construct.
Manufacturer(s)
DePuy Spine, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →