RecallWatchUSA

Home / Recalls / Medical Devices

DeRoyal Industries Inc recalls DeRoyal(R) SafeLiner Suction Canister Kit, SEMI-RIGID, 1000cc, REF 71-6518, NON-STERILE, Rx Only, general hospital

Recalled October 29, 2014 · FDA recall Z-0098-2015 · DeRoyal Industries

Hazard
These lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition. · FDA Class II
Units
5112 units
Sold at
Nationwide Distribution including MI, OH, IN, FL, WV, VA, MI, NC, and NY

Description

These lots of suction canisters and lids have an insufficient seal under low or intermittent suction condition.

Product
DeRoyal(R) SafeLiner Suction Canister Kit, SEMI-RIGID, 1000cc, REF 71-6518, NON-STERILE, Rx Only, general hospital
Manufacturer(s)
DeRoyal Industries Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov