DeRoyal Industries Inc recalls DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C

Recalled March 12, 2025 · FDA recall Z-1291-2025 · DeRoyal Industries

Hazard
Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care. · FDA Class II
Units
714 cases
Sold at
U.S.: IL, MN, NC, PA, and VA O.U.S.: N/A

Description

Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.

Product
DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement
Manufacturer(s)
DeRoyal Industries Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov