RecallWatchUSA

Home / Recalls / Medical Devices

Diagnostica Stago, Inc. recalls STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop a

Recalled September 3, 2025 · FDA recall Z-2473-2025 · Diagnostica Stago

Hazard
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications. · FDA Class II
Units
3,650 units (538 US)
Sold at
Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay, Peru, Republique Tcheque, Romania, Saudi Arabia, Serbia, South Africa, Switzerland, Taiwan, Thailand, Tunisia, Turquie, United Arab Emirates, Uruguay, Vietnam.

Description

Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.

Product
STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop analyzer capable of simultaneously performing clotting, chromogenic, and immunologic assays.
Manufacturer(s)
Diagnostica Stago, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Diagnostica Stago, Inc. recalls STA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop a | RecallWatchUSA