Diasorin Inc. recalls DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens

Recalled December 18, 2024 · FDA recall Z-0677-2025 · Diasorin

Hazard
Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result. · FDA Class II
Units
21905 units
Sold at
Worldwide

Description

Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.

Product
DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens
Manufacturer(s)
Diasorin Inc.