Draegar Medical Systems, Inc. recalls Draeger Medical Systems Infinity Acute Care System (M540) Catalog Numbers: MK31501/MK31701/MK31722
Recalled July 26, 2017 · FDA recall Z-2734-2017 · Draegar Medical Systems
Hazard
Cockpits with revision index 06 or higher that contain 4GB RAM modules may not annunciate audio or visual alarms on the Cockpit and Central Station. · FDA Class II
Units
37 US and 682 OUS
Sold at
AZ, NH, SD, TN, TX Foreign: Australia Austria Belgium Canada China Costa Rica France Germany Great Britian Hong Kong India Ireland Italy Japan Malaysia Namibia Netherlands Poland Portugal Russian Fed. South Africa South Korea Spain Sweden Switzerland Thailand Turkey United Kingdom
Description
Cockpits with revision index 06 or higher that contain 4GB RAM modules may not annunciate audio or visual alarms on the Cockpit and Central Station.
Product
Draeger Medical Systems Infinity Acute Care System (M540) Catalog Numbers: MK31501/MK31701/MK31722
Manufacturer(s)
Draegar Medical Systems, Inc.