Draegar Medical Systems, Inc. recalls INFINITY DUAL HEMO MCable Pod
Recalled March 21, 2018 · FDA recall Z-0942-2018 · Draegar Medical Systems
Hazard
Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dr¿ger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure. · FDA Class II
Units
2,760
Sold at
Nationally
Description
Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Dr¿ger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an incorrect measured value of invasive blood pressure.
Product
INFINITY DUAL HEMO MCable Pod
Manufacturer(s)
Draegar Medical Systems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →