Home / Recalls / Medical Devices
Draeger Medical, Inc. recalls Anesthesia Set Coax 180, MP03384
Recalled May 24, 2023 · FDA recall Z-1458-2023 · Draeger Medical
Hazard
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces. · FDA Class I
Units
20 units
Sold at
Domestic: US Nationwide Distribution.
Description
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.
Product
Anesthesia Set Coax 180, MP03384
Manufacturer(s)
Draeger Medical, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →