Home / Recalls / Medical Devices
Draeger Medical, Inc. recalls Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.
Recalled October 20, 2021 · FDA recall Z-0109-2022 · Draeger Medical
Hazard
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds · FDA Class II
Units
2,169 units
Sold at
US Nationwide
Description
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Product
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Manufacturer(s)
Draeger Medical, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →