Draeger Medical, Inc. recalls Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal

Recalled October 20, 2021 · FDA recall Z-0108-2022 · Draeger Medical

Hazard
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds · FDA Class II
Units
6,797 units
Sold at
US Nationwide

Description

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Product
Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400
Manufacturer(s)
Draeger Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov