RecallWatchUSA

Home / Recalls / Medical Devices

Draeger Medical, Inc. recalls Seattle PAP Plus, MP02260

Recalled May 24, 2023 · FDA recall Z-1439-2023 · Draeger Medical

Hazard
Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces. · FDA Class I
Units
32,050 untis
Sold at
Domestic: US Nationwide Distribution.

Description

Glued connections of the breathing circuit can become loose before or during ventilation resulting in partial or complete detachment of pieces.

Product
Seattle PAP Plus, MP02260
Manufacturer(s)
Draeger Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Draeger Medical, Inc. recalls Seattle PAP Plus, MP02260 | RecallWatchUSA