Draeger Medical Systems, Inc. recalls Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833

Recalled April 16, 2014 · FDA recall Z-1440-2014 · Draeger Medical Systems

Hazard
The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged. · FDA Class I
Units
3
Sold at
USA Nationwide Distribution in the states of WI and NY.

Description

The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.

Product
Draeger Oxylog 3000 plus ventilators, emergency and transport ventilator. Part number: 5704833
Manufacturer(s)
Draeger Medical Systems, Inc.