RecallWatchUSA

Home / Recalls / Medical Devices

Ebi, Llc recalls Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Inte

Recalled December 31, 2014 · FDA recall Z-0874-2015 · Ebi

Hazard
A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation. · FDA Class II
Units
193 units
Sold at
US Nationwide Distribution.

Description

A design deficiency was discovered whereby the Drill/Awl Sleeve and Spring-Loaded Drill/Awl Sleeve instruments (the Sleeves) allow the mating Drill and Awl instruments to penetrate 1.0 mm beyond their intended depths during screw hole preparation.

Product
Solitaire-C Cervical Spacer System a) 14-531593 Drill/Awl Sleeve b) 14-531594 Spring-Loaded Drill/Awl Sleeve Intended to stabilize cervical spinal segment to promote fusion in order to restrict motion and decrease pain using bone graft with or without supplemental fixation.
Manufacturer(s)
Ebi, Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov