Edwards Lifesciences, LLC recalls Edwards SAPIEN 3 Ultra Transcatheter Heart Valve System (SAPIEN 3 Ultra Delivery System) (26mm) 9630TF26 UDI:006901032

Recalled August 28, 2019 · FDA recall Z-2226-2019 · Edwards Lifesciences

Hazard
The action has been initiated to address reports of burst balloons which have resulted in significant difficulty retrieving the device into the sheath and withdrawing the system from the patient during procedures · FDA Class I
Units
2579 Units
Sold at
US: VA, CA, GA, MA, MD, NY, OR, WA, MN, TN, AL, KS OUS: United Kingdom, Switzerland, Sweden, Spain, Portugal, Poland, Norway, Netherlands, Italy, Ireland, Germany, Finland, Denmark, Canary Islands, Canada, Austria

Description

The action has been initiated to address reports of burst balloons which have resulted in significant difficulty retrieving the device into the sheath and withdrawing the system from the patient during procedures

Product
Edwards SAPIEN 3 Ultra Transcatheter Heart Valve System (SAPIEN 3 Ultra Delivery System) (26mm) 9630TF26 UDI:00690103201253
Manufacturer(s)
Edwards Lifesciences, LLC